SW
PROTOCOL
SAMPLING,
ANALYSIS, AND REPORTING RESULTS
Samples are enumerated for Escherichia coli (E. coli)
in the field by use of the modified mTEC method. See Escherichia coli using the modified
mTEC method. For each sample, a
series of volumes are plated to obtain the ideal plate count of 20-80 colonies
per plate. A list of recommended and required quality-control samples for
bacteria is in Quality-control
samples for NAWQA microbiology.
Refer to the USGS National
Field Manual, Biological Indicators for detailed instructions on sample
collection and analysis, parts of which are included below.
Equipment and supplies. Refer to the
list of Equipment and
Supplies. For rinsing the filtration
equipment and making dilutions, use “saline” buffer made by Ocala (Q373BACT,
250 mL volume and Q421BACT, 99-mL volume) or sterile, buffered water, PO4/MgCl2 made
by Ocala (Q453BACT, 250-mL volume) and the 47-mm, 0.45-mm sterile membrane filters
(OCALA, Q353BACT). Modified mTEC medium is purchased as dehydrated agar from Becton-Dickinson
(BD), catalog number 214884.
Clean and sterilize equipment. Wash and
gently scrub equipment (filter funnels, graduated cylinders, caps and nozzles,
1-L sample bottles) with 0.1 percent liquinox.
Thoroughly rinse the scrubbed items with warm tap water until there is
no sign of any detergent. Then, thoroughly rinse all equipment with deionized
water (DIW). After cleaning, cover the open ends with aluminum foil or kraft paper and autoclave for 15 minutes at 121o
C and 15 lb/in2. Use heat-sterilizing tape to ensure proper
performance of the autoclave.
Prepare a separate set
of sterile equipment (filter funnels, graduated cylinders, caps and nozzles,
1-L sample bottles) for microbiological sampling at each site.
Heat incubators. Heat two
incubators or chambers—one at 35°C and another at 44.5°C. It takes about 2
hours for the incubators to reach desired temperatures
Prepare modified mTEC plates. Prepare mixture as directed on the dehydrated
medium bottle (45.6 grams agar to 1 liter deionized
water). Heat to
boiling with a stirring rod on a hot plate. Add approximately 100-mL of agar to each
dilution bottle. Autoclave
for 15 minutes. Store dilution bottles at 4°C
for up to six months. Melt agar
either for 5 minutes in an autoclave or in a beaker of boiling water on a
hotplate until melted. Be sure to loosen
the cap prior to melting. Allow the agar
to cool until it is warm (50-60°C). Gently mix the agar and pour
plates. Store plates
upside down in a plastic bag at 1-4°C. Plates
are good for two weeks in the refrigerator.
You must keep the plates chilled at all times.
Collect bacteria sample. Because surface-water sampling for NAWQA Source
Water Quality Assessments is done near the intake to drinking-water plants,
samples for E. coli will be collected using the hand-dip (grab) sampling
method. See Stream water sample
collection. Do not obtain the E. coli
sample from churn or cone-splitters, as these devices are not sterile. Instead, use a separate, sterile, 1-L bottle
to collect the sample for E. coli.
Be sure to leave at least 2 inches of headspace in the bottle.
After collection,
immediately chill samples in a cooler with ice or in a refrigerator at 1 to 4°C. SAMPLES
NEED TO BE PLATED FOR BACTERIA WITHIN 6 HOURS OF COLLECTION.
Plating samples for bacteria [Refer to the USGS
National Field Manual, Biological Indicators for detailed
membrane-filtration procedures.]
Label your plates and gather forms. Decide on
appropriate sample volumes to obtain at least one plate in the ideal-count
range of 20-80 colonies. If the
bacterial water-quality of the stream is unknown, plate the following sample
volumes: 100, 30, 10, 3, 1, 0.3, and 0.1
mL. For less
polluted waters, the three smallest volumes may be omitted. For more polluted waters, the two highest
volumes may be omitted. As you become
more familiar with the water-quality of the stream, the number of sample
volumes plated can be reduced.
On the lid of the plate, record the site, date,
sample-collection time, and the volume (or FB for filter blank and PB for
procedure blank). Use the SW sample and positive
control report form for the regular sample and the positive control
culture, and the SW sample
and replicate report form for
the regular and associated replicate samples. (See Quality control section
below for information on positive controls and replicates).
Assemble membrane filtration apparatus and gather
supplies. Place the filter funnel in
a manifold attached to a peristaltic pump. Use the same filter funnel to plate
the series of volumes for one sample. Use a different sterile filter funnel for
each surface-water sample.
Prepare sample dilutions (if needed). To accurately
deliver sample volumes less than 1.0 mL, you must
prepare dilutions of the sample. For
example, for more polluted waters, plate 30- and 10-mL volumes of a 1/100
dilution (these equal 0.3 and 0.1 mL of actual sample
volumes, respectively). To prepare a
1/100 dilution, label a 99-mL bottle of saline buffer as the 1/100 dilution and
loosen the bottle cap. Place the pipet bulb on a 1-mL pipet (kept
sterile in the original wrapper). Shake
the sample at least 20 times. Immediately
remove a 1-mL aliquot from the center of the sample and transfer it to the
dilution bottle. If the bottle sits on
the counter for more than a few seconds before removing the 1-mL aliquot, you
need to shake the bottle again—bacteria settle quickly to the bottom of the
sample bottle.
Filter the filter blank. First,
a filter blank (FB), which tests the sterility of buffered water and filtration
equipment, is filtered. This is done by filtering a 50-100 mL aliquot of sterile buffered water. Use the gradations on the side of the filter
funnel to measure a 50-100 mL aliquot of buffered
water.
If contamination from a FB is found, results are suspect and may be qualified or not reported. For criteria, see Quality-control samples for NAWQA microbiology.
Filter the sample. Filter
the sample in order of smallest to largest sample volume. Use a graduated cylinder to
measure volumes greater than 10 mL. You can also use
a 25-mL pipet to measure volumes of 30 mL (go to –5 mL on the pipet). Use a 10-mL pipet to measure volumes from 2 to 10 mL.
Use a 1-mL pipet to measure 1-mL samples
volumes.
FOR VOLUMES LESS THAN 10 ML,
PUT ABOUT 10-ML OF BUFFER INTO THE FUNNEL BEFORE ADDING THE SAMPLE. This ensures even
distribution of the sample across the membrane.
FOR ALL VOLUMES, RINSE THE
FILTER FUNNEL WITH BUFFER AFTER FILTERING THE SAMPLE AND BEFORE REMOVING
THE FILTER.
This ensures that the bacteria are washed off the side of the funnel
adequately.
BE SURE TO THOROUGHLY SHAKE THE SAMPLE BEFORE MEASURING
THE VOLUME. If you don’t, you will not get a
representative subsample for plating. If the bottle sits on the counter for more
than a few seconds before removing the sample aliquot for plating, you need to
shake the bottle again.
The three bolded and capitalized steps are VERY
IMPORTANT!
Filter the procedure blank (these are included after
every tenth sample). The
procedure blank (PB) tests the adequacy of the rinsing procedure. This is done by filtering a 50-100 mL aliquot of sterile buffered water after
filtering the sample, using the same
filter funnel.
Incubate plates.
Invert the plates and incubate for 35°C for 2 hours.
Transfer all plates to an incubator set at 44.5°C for 22 hours. Fill out the report form.
Count
E. coli. Count all red or magenta
colonies as E. coli. Using a
dissecting microscope or portable illuminated magnifier to count colonies is
highly recommended.
Disposal. After counting, place all plates in an autoclave bag and autoclave for 30 minutes at 121oC. Dispose of the bag in the trash.
Calculate the colonies per 100 milliliters
(col/100 mL) as described in the USGS National
Field Manual, Biological Indicators.
![]()
The results from the membrane-filtration
tests are entered into NWIS using the following parameter code:
90902, E. coli on modified mTEC agar, colonies
per 100 mL
If counts are not within the ideal range of
20-80 colonies, enter a remark code of “E” and a value qualifier of “k”.
If the filter has a colony count greater than
80 and the colonies are distinct enough to count, report the number and enter a
remark code of “E” and a value qualifier of “k”. If the plate is too numerous
to count (TNTC) and the colonies are not discernable, report a minimum
estimated value. Assume a count of the maximum range of colonies (80 for E.
coli on modified mTEC agar) for the smallest sample volume filtered and
report the result as greater than the calculated value per 100 mL. For example, assume 80 colonies on the 1 mL volume filtered to get a result of >8,000 colonies
per 100 mL, with a value qualifier of “k”.
·
Check the incubators daily to ensure they are
operating at the proper temperature. Use an external digital thermometer that
enables you to take a temperature reading with the incubator closed (see
equipment list). Record the results in a
QA/QC logbook. Criteria for acceptance
are 35 ± 0.5°C or
44.5 ± 0.2°C.
·
The gray incubators purchased from Millipore
sometimes respond to a power surge by shutting off. A good practice is to periodically monitor
the incubator temperature with a HOBO temperature recorded for 48 hours throughout
field and laboratory use.
·
Use heat-indicating tape with each autoclave run to ensure it is
operating properly.
·
Test the autoclave with biological indicators quarterly or more often
during heavy use (you can order these from a scientific supplier—Barnstead No.
AY759X7).
·
Before processing samples in the field vehicle, wipe down the area with
dilute (50/50) isopropyl alcohol.
·
After counting, turn the plate 180° and ensure the second count is within 5 percent of the first count.
·
Have a second analyst check calculations for errors.
Quality control
Collect and analyze quality-control samples
at the required frequency. See Quality-control samples for
NAWQA microbiology.
The following QC is required:
·
Filter blanks—a 50-100 mL aliquot of sterile buffered water
filtered before the sample.
Filter blanks are used to
qualify associated environmental sample.
Filter blanks results are not stored in NWIS.
·
Procedure blanks—a 50-100 mL aliquot of sterile buffered water
filtered after the sample. Include
procedure blanks if the analyst is inexperienced in membrane filtration and
(or) if water is suspected to have high concentrations of target colonies. If the procedure blank is positive, this
indicates that rinsing techniques may be inadequate. Procedure blanks are used to qualify
associated environmental sample.
Procedure blank results are not stored in NWIS.
·
Positive controls of E. coli are required for
surface-water sampling. They are
especially important when first using the method—to assess if a good dilution
series is obtained and whether the analyst can correctly identify E. coli. Positive controls for surface water can be
obtained from the Ohio District Microbiology Laboratory (ODML). The controls come in 100-mL aliquots in
dilution bottles, and are plated in the field using 30-, 10-, 3-, and 1-mL
sample volumes. The positive-control
bottles must be plated the same day they are received from the ODML. Use the SW sample and positive
control report form to record results.
Instructions for analyzing the positive controls are in SW control instructions.
Positive control results
are not stored in NWIS. Send your
results to the Ohio District (amgbrady@usgs.gov).
·
Negative-controls are not recommended for surface-water sampling because
the agar is very stable and target colonies are easily discerned from
non-target colonies.
·
Field blanks—Pour about 250 mL of
sterile buffered water into a clean and sterile sample bottle. Process as you would any other sample.
Field blank results are
stored in QADATA with the following codes:
o Sample-medium code is “Q”
for quality control
o Sample-type code is “2”
for blank
o Parameter code 99100
(blank-solution type), is “70” for sterile buffered water PO4/MgCl2 or “60” for
sterile saline buffered water (lot number is not required)
o Parameter code 99101
(source of blank water) is “80” for
o Parameter code 99102
(blank-sample type) is “100” for field
·
Replicates—The replicate is collected as a split
sequential grab sample. This means that
two bottles are collected and each bottle is plated twice. Use the SW sample and replicate
report form to record results of the regular and replicate
samples.
Enter the 1st sample in
QWDATA with the following codes:
o Sample medium is “9” for
surface water
o Sample type is “7” for
replicate
o Parameter code 99111 (QC
sample associated with this environmental sample) is “30” for replicate sample,
or “100” if more than 1 type of field QC sample is collected
o Parameter code 99105 (replicate type) is “50” for split sequential
Enter the 2nd
to 4th samples in QADATA with the following codes:
o Sample medium is “R” for
surface water
o Sample type is “7” for
replicate
o Parameter code 99105 (replicate type) is “50” for split sequential
NOTE: Offset sample times so that splits of the
same grab sample are closer in time than splits of different grab samples. For
example: 1300, 1301, 1305, and 1306 where 1300 is the 1st environmental sample
(grab sample 1, split 1), 1301 is the 2nd sample (grab sample 1, split 2), 1305
is the 3rd sample (grab sample 2, split 1) and 1306 is the 4th
sample (grab sample 2, split 2).
Updated February 2004