Ohio Water Microbiology Lab
Research Topics
Internal Information
Ohio Water Science Center
USGS In Your State
USGS Water Science Centers are located in each state.
|
OWML: QA/QC
Quality Assurance/Quality Control Manual: Ohio Water Microbiology
Laboratory
Laboratory water
The OWML has three types of laboratory water:
-
Type III deionized water (“deionized water”) produced from City of Columbus
tap water for general laboratory use. The deionized water unit and tap are
stored in the warehouse. The system is described in Francy and others
(1998). The vendor changes the cation and anion columns, moves forward the
standby mixed-bed column, installs a new standby tank, and changes the
carbon filter when the red service light illuminates. Maintenance checks
are recorded in the LIMS.
-
Reagent-grade water produced using a Millipore MilliQ system (“MilliQ
water”). Deionized water is used as source water for the MilliQ system.
Reagent water is used for cultivation media and additives (mTEC, MI, mEI,
antibiotic stocks, and others) as well as for preparation of reagents for
sensitive procedures (elutions, PCR, hybridization, and others). The MilliQ
cartridges are changed by OWML laboratory personnel when the service light
blinks and the display message reads “EXCH. CARTRIDGES.” Indicate the
date of cartridge change in the LIMS.
-
Deionized water is stored in a laboratory carboy (“stored water”) and used
for rinsing of dishware and other supplies.
A variety of quality-control checks are routinely done on the three types of
water and may differ depending on the type of water. Acceptance criteria are
listed in table 2. For deionized water, two levels of acceptance criteria are
listed—(1) a warning level wherein the system is inspected and constituents are
retested and (2) a shut-down level. For MilliQ water, only a shut-down level is
listed in table 2. For stored water, if criteria are not met, the container is
cleaned out, refilled, and retested.
-
Quarterly checks of specific
conductance and turbidity are done on all three types of water and recorded
in the LIMS.
Instructions for
performing this check are in the back of the equipment QA/QC logbook.
-
Quarterly checks of bacterial growth
are done on the MilliQ water and recorded in the LIMS. Instructions for
performing this check are in the back of the equipment QA/QC logbook.
-
A blank of deionized water is submitted to the National Water Quality
Laboratory (NWQL) annually and analyzed for low level nutrients (Schedule
1217), and total-organic carbon (Labcode 114), and the results are recorded
in the LIMS.
We no
longer analyze a blank for trace elements and low-level major ions because
the need for these low-level analyses is project specific.
-
The stored deionized water carboy is
to be emptied completely and cleaned with Liquinox and water every other
week. Record cleanings in the LIMS.
Table 2. Acceptance criteria for laboratory water quality-assurance
checks[Adopted from USEPA (1978), APHA (1998), and ASTM (1999); NA is not
applicable; constituents highlighted in gray are no longer required tests]
|
|
DEIONIZED |
MILLIQ |
STORED |
|
ACTION |
warning |
shut down |
shut down |
clean and refill |
|
Specific conductance (ms/cm) |
3 |
5 |
2 |
3 |
|
Turbidity |
1 |
5 |
1 |
1 |
|
Heterotrophic plate count (colonies/mL) |
NA |
NA |
<1 |
NA |
|
Total organic carbon (mg/L) |
0.2 |
10 |
NA |
NA |
|
Sodium (mg/L) |
0.1 |
1 |
NA |
NA |
|
Nutrients individual (mg/L) |
0.1 |
1 |
NA |
NA |
|
Heavy metals, individual (Cd, Cr, Cu, Ni, Pb, Zn) (mg/L) |
1 |
10 |
NA |
NA |
|
Other trace elements (mg/L) |
3 |
50 |
NA |
NA |
|